Solvent Management  
Solvent Conditioning Vacuum degas, three solvents plus purge solvent
Typical Operating Flow Rate Range 0.050 to 5.000 mL/min in 0.001-mL/min increments
Programmable Flow Rate Range .000 and 0.010 to 10.000 mL/min in 0.001-mL/min increments
Compressibility Compensation Automatic and continuous
System Delay Volume <400 µL, independent of backpressure, 1 mL/min
Plunger Seal Wash Integral, active, programmable
Gradient Profiles: Eleven (11) gradient curves [including linear, step (2 curves), concave (4 curves), and convex (4 curves)]
Gradient Profiles Eleven (11) gradient curves [including linear, step (2 curves), concave (4 curves), and convex (4 curves)]
Dry Prime/Wet Prime Automatic, front panel control
Flow Ramping Time (0.01 to 30.00 min in 0.01-min increments) to reach maximum flow rate
Maximum Operating Pressure 5000 psi (345 bar) at 0.010 to 3.000 mL/min; Programmable upper and lower limits
Composition Range 0.0 to 100.0% in 0.1% increments
Composition Accuracy ±0.5% absolute, independent of backpressure (Proportioning Valve Pair Test [degassed methanol: methanol/propylparaben, 2 mL/min, 257 nm])
A Composition Precision
 
<=0.15% RSD or<=0.02 min SD, whichever is greater, based on retention time. (Degassed 60% methanol:40% water Dial-a-Mix, 1 mL/min, six replicates, phenone mix, 257 nm)
Flow Precision <=0.075% RSD or <=0.02 min SD, whichever is greater, based on retention time (N = 6) or volumetric measures (0.200 to 5.000 mL/min), isocratic premix
Flow Accuracy <=1% or 10 µL/min, whichever is greater, (0.200 to 5.000 mL/ min), degassed methanol, at 600 psi backpressure

Sample Management

 

umber of Sample Plates: Total of four plates: 
– Microtiter plate – 96- or 384-well 
– Vial plate – 2.0-mL vial (48) 
– Tube plate – 0.5-mL microcentrifuge tube (48) or 1.5-mL microcentrifuge tube (24) 
– Open access plate – 2.0-mL vial (24 or 48)
Sample Temperature Control: 4 to 40 °C, programmable in 1 °C increments
Maximum Sample Capacity 1536 in four 384-well plates
Number of Sample Injections: 1 to 99 injections per sample
Sample Delivery Precision:  <0.3% RSD, full loop 50 µL. (Default wash/purge conditions, degassed 60% methanol: 40% water Dial-a-Mix, 1 mL/min, six replicates, paraben mix, 257 nm)
Sample Volume Linearity  >0.999% correlation, 5 to 20 µL, partial fill in 50-µL loop, 3-µL pre- and post-sample air gaps (Default wash/purge conditions, degassed 60% methanol:40% water Dial-a-Mix, 1 mL/min, six replicates, paraben mix, 257 nm)
Needle Wash Solvents Two: wash (strong solvent) and purge (sample- compatible) solvent A
Sample Carryover <0.01% or <2.5 nL (default wash/purge volumes), whichever is greater. (Blank injection, 25 µL, 60% methanol:40% water, after six replicates, 25 µL propylparaben, 1 g/L, 257 nm)

Waters 474 scanning fluorescence detector

Condition Specification
Monochromator Holographic concave diffraction grating (for both excitation and emission)
Light source 150 W Xenon lamp
Wavelength range 200 to 700 nm (for excitation and emission). Can be set from 0 to 900
Spectrum band width Excitation: 18 nm
Emission: 10, 18, or 40 nm (set on keypad)
Wavelength accuracy ±2 nm
Wavelength repeatability ±0.3 nm
Detector Excitation: Photodiode
Emission: Photomultiplier tube
Second order filter Automatically inserted into light path when Ex > 400 Nm
Flow cell 16 µL (standard) or 5 µL
Maximum flow cell pressure 16 µL: 145 psi (10 bar)
5 µL: 500 psi (34.5 bar)
Wetted surfaces Synthetic silica, fluororesin, PEEK, stainless steel
SXS-316
Sensitivity S/N for the Raman peak of water is 320 or above, where Ex λ = 350 nm and time constant is 1.5 sec, 16 µL cell (Instruments shipped prior to June 1997, S/N for the Raman peak of water is 200 or above)
Attenuation range (Recorder output) Ten steps, including 1, 2, 4, 8, 16, 32, 64, 128, 256, S
Gain range 1, 10, 100, 1000
Response Fast, Standard, Slow, and digital filter

Signal processing Digital processing by A/D and D/A converters (with  surrounding temperature compensation circuits)
Outputs • Recorder: 10 mV full-scale (can be inverted)
• Integrator: 1 V full-scale
• Marker output
• Leak output
Inputs • Marker input
• Auto Zero input
• Program reset
Program functions Time program with following variables:
• Wavelength change (excitation and emission)
• Gain
• Attenuation
• Wavelength scan
• Ex, Em Spectra
Wavelength scan functions Manual and time program parameters:
• Time
• Interval
• Amplitude
• Memory location
• Ex, Em Spectra
Spectrum storage Ten locations for both emission and excitation spectra
Self-test diagnostics • Memory (ROM and RAM)
• DC power
• Excitation energy decrease
• Leak in cell
• Lamp use time
Lamp use timer Internal timer for accumulating the total hours on the
Xenon lamp
Dimensions Width: 11.81 inches (300 mm)
Height: 5.90 inches (150 mm)
Depth: 18.11 inches (460 mm)
Weight 41.8 pounds (19 kg)
Power requirement 115 VAC (90 to 132 V), 50/60 Hz
230 VAC (187 to 264 V), 50/60 Hz
Power consumption 300 W maximum
Ambient temperature 10 °C to 32 °C (50 °F to 90 °F) operation
10 °C to 60 °C (50 °F to 140 °F) storage
Relative humidity 20 to 80%, non-condensing

This refurbished Waters Alliance 2795 HPLC system comes with Waters 474 Scanning Fluorescence  Detector.
 
We have other Waters HPLC detectors as option.
 
    * Waters 486 Tunable Detector
    * Waters 2487 Dual wavelength Detector
    * Waters 996 PDA Detector
    * Waters 2996 PDA Detector
 
This refurbished Waters Alliance 2795 HPLC system with Waters 474 scanning fluorescence detector is in great condition. It has been tested at SpectraLab Scientific.
 
We could sell them separately upon request.
 
A 90 day warranty is included. Multiple available.

SpectraLab Scientific Inc
Established in 2003,

SpectraLab Scientific is a global leader of Refurbished analytical equipment. We specialize in reconditioning equipment, integrating the best available technologies from leading manufacturers across the industry. We are committed to quality and service, delivering strong customer satisfaction. We stock over 10,000 pre-owned lab equipment, components, and mix-and-match parts for your systems.

Select the fields to be shown. Others will be hidden. Drag and drop to rearrange the order.
  • Image
  • SKU
  • Rating
  • Price
  • Stock
  • Availability
  • Add to cart
  • Description
  • Content
  • Weight
  • Dimensions
  • Additional information
Click outside to hide the comparison bar
Compare