This refurbished HP 1050 Quat Pump is in great condition. It has been tested at SpectraLab Scientific.

A 90 day warranty is included.

Hydraulic System Dual-pistons in-series with proprietary servo-controlled
variable stroke drive floating pistons and active inlet valve.
Flow Range Setpoint from 0.001 to 9.999 ml/min in 0.001 ml/min increments.
Piston Displacement 20 to 100 µl, automatic matched to flow rate or user-selectable.
Flow Precision <0.3% RSD (typically <0.15%) based on retention time at 0.5 ml/min and 2.5 ml/min.
Pressure Operating range from 0-400 bar (5880 psi) up to 5 ml/min; from 0-200 bar (2950 psi) up to 10 ml/min. Display in bar, psi or MPa
Pressure Pulsation <2% amplitude (typically <1%), 1 ml/min isopropanol at all pressures >10 bar
Compressibility Compensation User-selectable, based on mobile phase compressibility
Recommended pH Range 2.3 to 12.5 (stainless steel version). Solvents with pH below 2.3 should not contain acids which attack stainless steel. 1.0 to 14 (TI series).
Gradient Formation Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve. Delay Volume
900 to 1100 µl dependent on back pressure.
Composition Range 0 to 100% in 0.1% increments from four independent channels.
Composition Precision ±0.25% absolute (typically ±0.15%) peak to peak, binary mixed water/acetonitrile from 0.5 ml/min to 5.0 ml/min without mixer.
Solvent Preparation Four 1 liter bottles each with individually-regulated Helium sparger, cap and filter.
Oven Temperature Range Ambient +5°C to ambient +60°C in 0.1°C increments. Display in °C, °F or K.
Oven Temperature Stability ±0.15°C
Oven Capacity Two 25-cm or three 20-cm columns.
Control Integrated keyboard with function keys; parameter editing during run possible; keyboard lock; optional control by PC
Parameters Flow rate, compressibility, stroke volume, upper and lower pressure limits, 2 external contacts; %B, %C, %D (for quaternary pump). Time-programmable Parameters: Flow rate, upper pressure limit, external contacts; %B, %C, %D.
Methods Battery-backed storage of up to 10 methods. Automatic start up and shut down methods. Editing of stored methods possible in run.
Analog Output For pressure monitoring, 2 mV/bar.
Communications Outputs: ready signal and two external outputs (one 24 V relay and one 30V (AC/DC) contact closure, both with 0.25
A. In-and outputs: start, stop and shut down signals. Optional interface for GPIB and RS-232C.
Safety Aids Extensive diagnostics, error detection and display via front-panel LED’s and status logbook. User-definable shutdown method activated in case of error. Leak detection and safe leak handling. Low voltages in major maintenance areas. Column pressure protection with maximum rate of pressure change of <20 bar/sec after a setpoint change.
Environment 4°C to 55°C (constant temperature) with <95% humidity (non-condensing).
Power Requirements Line voltage: 100-120 or 220-240 VAC ±10%

SpectraLab Scientific Inc
Established in 2003,

SpectraLab Scientific is a global leader of Refurbished analytical equipment. We specialize in reconditioning equipment, integrating the best available technologies from leading manufacturers across the industry. We are committed to quality and service, delivering strong customer satisfaction. We stock over 10,000 pre-owned lab equipment, components, and mix-and-match parts for your systems.

Select the fields to be shown. Others will be hidden. Drag and drop to rearrange the order.
  • Image
  • SKU
  • Rating
  • Price
  • Stock
  • Availability
  • Add to cart
  • Description
  • Content
  • Weight
  • Dimensions
  • Additional information
Click outside to hide the comparison bar
Compare