Shimadzu Gas Chromatograph System High Performance and Expandability Merged at a higher level

High Performance

Improved design and innovative technology for all of our injectors, detectors and flow controllers equal or surpass our GC-2010 the high-end technology leader

Easy Operation

Large LCD, all digital gases control and auto-diagnostics inherited from the GC-

2010 – “The Most advanced, easy-to-use interface.

Flexibility

Use any column types for any analysis. Packed or capillary columns give you

the freedom to choose the best technique for your measurement. Fully integrated multiple valve systems are made simple for optimum performance for SystemGC custom GC products.

Column Oven

 

Temperature range

(Ambient + 10°C) ~ 400°C (using liquid CO2 gas*: -50°C ~400°C)

Dimensions

250 (W) × 360 (H) × 175 (D) mm

Oven capacity

15.8L

Temperature accuracy

Set value (K) ± 1% (calibration at 0.01°C increments)

Temperature deviation

2°C max. (on 200mm dia. circumference 30mm from rear)

Temperature variation coefficient

0.01°C/°C

Temperature program steps

Up to 20 (cooling program possible)

Programmed rate setting range

-250°C ~ 250°C/min

Total time for all steps

9999.99 minutes max

Linear heating range

30°C/min up to 150°C

20°C/min up to 250°C

10°C/min up to 380°C

7°C/min up to 400°C (at 25°C ambient temperature)

Cooling rate

300°C ~ 50°C in 6 min max. (at 25°C ambient temperature)

Columns accepted

Capillary columns: 2

Packed columns for GC14B: 4 (Glass columns: 2)

Sample Injection Unit

 

Temperature range

Up to 400°C

Heating settings

1°C steps

No. of units installed simultaneously

Up to 3 units

Sample injection unit types

Dual packed, single packed, split/splitless, direct, direct (AMC)

Carrier Gas Flow Controller

 

For Packed / Dual

Flow rate setting range

0 ~ 100mL/min

Programmable steps

7

Programmed rate setting range

400 ~ 400mL/min

Correction function

Maintains column ­flow rate during column oven heating

For Capillary Split/Splitless, Direct

 

(Split/splitless injection mode)

 

Pressure setting range

0 ~ 970kPa

Programmable steps

7 (pressure-decreasing program possible)

Programmed rate setting range

-400 ~ 400kPa/min

Split ratio setting range

0 ~ 9999.9

Total fl­ow rate setting range

0 ~ 1200mL/min

Correction function

Maintains column average linear velocity during column oven heating (for capillary only)

(Pressure mode direct injection)

 

Pressure setting range

0 ~ 970kPa/min

Programmable steps

7

Programmed rate setting range

-400 ~ 400kPa/min

Correction function

Maintains column average linear velocity during column oven heating

(Flow-rate mode direct injection)

 

Flow rate setting range

0 ~ 1200mL/min

Programmable steps

7

Programmed rate setting range

-400 ~ 400mL/min

For Single Packed, Direct (AMC)

 

Flow rate setting range

0~100mL/min

Correction function

Maintenance column ­flow rate during column oven heating

Detectors

 

Temperature range

400°C max. (FID, TCD)

Temperature setting

1°C steps

No. of units installed simultaneously

Up to 4 units (restricted depending on detector type)

Flame Ionization Detector (FID)

 

System

Dual ­flow rate differential system

Temperature range

400°C max.

Minimum detected quantity

3pgC/s (dodecane)

Dynamic range

107

Nozzle

Quartz glass Standard: for packed, Option: for capillary

Thermal Conductivity Detector (TCD)

 

System

Dual flow rate differential system

Temperature range

400°C max.

Dynamic range

105

Sensitivity

40,000mV · mL/mg (built-in pre-amplifier, with 10 × application)

System

Fixed current system using 63Ni370MBq radiation source

Temperature range

400°C max.

Minimum detected quantity

0.1pg/s (γ-BHC)

Dynamic range

104

Flame Photometric Detector (FPD)

 

Temperature range

350°C max.

Dynamic range

P: 104 S: 103

Minimum detected quantity

P: 0.5pgP/s (tributyl phosphate)

S: 8pgS/s (dodecane thiol)

Flame Thermionic Detector (FTD)

(Two types, one for capillary and one for packed. The speci‑cation are the same.)

Temperature range

400°C max.

Dynamic range

N: 103 P: 103

Minimum detected quantity

N: 0.4pgN/s (azobenzene)

P: 0.05pgP/s (malathion)

SpectraLab Scientific Inc
Established in 2003,

SpectraLab Scientific is a global leader of Refurbished analytical equipment. We specialize in reconditioning equipment, integrating the best available technologies from leading manufacturers across the industry. We are committed to quality and service, delivering strong customer satisfaction. We stock over 10,000 pre-owned lab equipment, components, and mix-and-match parts for your systems.

Select the fields to be shown. Others will be hidden. Drag and drop to rearrange the order.
  • Image
  • SKU
  • Rating
  • Price
  • Stock
  • Availability
  • Add to cart
  • Description
  • Content
  • Weight
  • Dimensions
  • Additional information
Click outside to hide the comparison bar
Compare