Solvent Delivery Pump (Degasser/mixer)

Specification  
Construction Quaternary low pressure gradient
Pump type Series double-plunger system (primary 10µ L, secondary 5µ L)
Solvent delivery system Low flow rate delivery
Low flow rate delivery Flow rate setting

range

0.001µ 5 mL/minµ 1.0µ 35.0MPaµ
Flow rate accuracy (1% or (2µ L/min, below whichever is greater (1mL/min, 3.0 ~ 10.0MPa, water,

room temperature fixed at 20(C, except during gradient solvent delivery)

(2% or (2µ L/min, below whichever is greater, except under above conditions

(except in gradient solvent delivery)

Flow rate precision 0.75%RSD or 0.02 min SD below whichever is greater

(0.05~2mL/min, 1.0 ~ 35.0MPa, methanol, room temperature fixed at 20(C)

Pressure limiter action Maximum value / minimum value setting possible
Plunger, seal Plunger: sapphire, seal: GFP
Check valves 2 valves: inlet side is dynamic inlet valve (DIV)
Solvent switching 4 solutions
Suction filter 10µm
Line filter 2µm
Drain valve There is also an auto-drain manual valve, which uses a rotary valve injectionsyringe.
Pulsation Less that 3MPa (3kg/cm2) (1mL/min of water, 100gk/cm2)
Pressure display accuracy 2% or 0.5MPa, whichever is greater
Pressure sensor 0-point auto correction Automatic zero adjustment upon instrument startup
Plunger rinse Auto rinse function (standard)
Purge function Auto Purge function, Manual Prime function
Gradient No. of solvents mixed 4 solutions
Mixing ratio range 0 ~ 100%, 0.1% steps
Concentration accuracy (0.5% (0.1 ~ 2mL/min, 1.0 ~ 20.0MPa, water and aqueous acetone mixture)
Concentration precision (0.1% (0.1 ~ 2mL/min, 1.0 ~ 20.0MPa, water and aqueous acetone mixture)
Mixer Bowl mixer (270µ L)
Leak capacity 800uL
Degasser Membrane online degasser, 5 lines (4 mobile phase, 1 rinse solution)
Safety mechanism Leak sensor Detection of leaks at degasser, solvent delivery pump, UV-VIS detector

Autosampler

Specification  
Injection system Full volume sample injection, variable injection volume (zero sample loss during injection)
Injection volume setting range 0.1 ~ 100µ L (standard), 1 ~ 2000µ L (option)
Number of samples processed Model without sample cooler

350 (w/1mL sample vials), 210 (w/1.5mL sample vials, 100 (w/ 4mL sample vials), 384

(w/ 4 96-well microtiter plates), 1536 (w/ 4 384-well microtiter plates)

384 (w/4 96-well deepwell plates)

Sample vials 1.5mL glass, 1.1mL glass, 1mL glass, 4mL glass, 0.3mL glass (w/plastic spacer), 0.3mL

glass (4mL vial storage type), 1mL plastic, 0.2mL plastic, 4mL plastic, 96-well microtiter

plate, 384-well microtiter plate, 96-well deepwell

Injection volume repeatability RSD<0.3% (10µ L injection)
Cross contamination Less than 0.01% (using caffeine analysis conditions)
Injection volume accuracy (1% (50µ L, n=6)
Injection time 15 seconds
Number of repeat injections 1 ~ 99 times/sample
Analysis time setting 0.01 ~ 9999.9 minutes (0.1 min steps)
Number of injection parameterSteps 202 steps max.
Sample suction speed 0.1 ~ 15µ L/s ( 0.1µ L/s steps)
Rinse solution suction speed 35 µ L/s
Applicable pressure 35.0 MPa max.
Sample No. BCD output Available (option)
Rinse solution degassing Standard

 

Column Oven

Specification  
Humidifying cooling  system Block heating system (mobile phase preheating function included)
Setting range 4 ~ 60°C(1°C steps)
4 ~ 60°C(1°C steps) (1°C (when using block surface temperature at specified location, and set temperature of 50°C);

(1°C (when temperature of mobile phase coming from preheater and set temperature are 40°C, at

1mL/min); 3(°C) at full flow rate range, temperature range

 

Temperature control precision 0.1°C
Temperature control range Room temperature -15°C ~ 60°C
Column oven interior Dimensions Two 25cm columns

 

Preision temperature control function Preheater function, 2 flow lines

Capacity: 9µ L (conventional), 4µ L (semi micro)

Safety mechanisms a) Maximum temperature setting to prevent overheating

b) Built-in temperature fuse

c) Built-in leak sensor

When leak sensor is activated, pump delivery, column oven termperature control, sample injection

Stop

Time Program function Controlled from system controller

UV-Vis detector

Specification  
Cell Temperature control built-in

Cell length: 10mm, capacity: 8µL

Light source Deuterium lamp
Wavelength range 190 ~ 600mm
Spectral bandwidth 8nm

 

Wavelength precision 1 nm
Wavelength repeatability 0.1 nm
Drift Less than 1 ( 10-4 AU/h(250nm, air in cell, fixed room temperature)
Noise level (0.5 ( 10-5 AU (corresponding to response 2s, 250nm, air in cell)
Linearity ~2.5Au
Auto wavelength accuracy check

Function

Check using low pressure (voltage) Hg lamp at 253.7nm, D2 lamp 656.1nm
Check using low pressure (voltage) Hg lamp at 253.7nm, D2 lamp 656.1nm Filter for absorbance check built-in
Response 10 step switching using time 0.05,0.1,0.5,1.0,1.5,2.0,3.0,6.0,8.0,10.0 seconds
Range From 0.0001-2.56 AUFS, 0.0001 AUFS step setting possible.
Zero correction Auto zero function
Polarity switching Possible

 

Cell recognition function Auto judgement of presence/absence of cell, cell type
Auto judgement of presence/absence of cell, cell type 2 wavelength monitoring of either 190 ~ 390nm or 371 ~ 600nm
Time Program Control from the controller possible
Output Integrator output 0.5, 1, 2, 4, 1.25, 2.5 Au/V step switching

System Controller

Specification  
Display Backlight equipped LCD display (320 ( 240 pixels), black and white
3.5 inch floppy disk drive Disk: 2HD, MS-DOS format
Input / output terminals External inputs: 2, external outputs: 2

Optical link: 1 (for Chromatopac)

RS-232C: 1 (for connection of PC)

Analog output: 2ch

Units that can be connected Recycle valve

BCD output option board (0-999 3-digit output possible)

Number of files Method files: 20

System preparation files: 10

Shutdown files: 2

Sequence files: 1

This refurbished SHIMADZU LC-2010A HPLC SYSTEM is in great condition. It has been tested at SpectraLab Scientific.

A 90 day warranty is included.

SpectraLab Scientific Inc
Established in 2003,

SpectraLab Scientific is a global leader of Refurbished analytical equipment. We specialize in reconditioning equipment, integrating the best available technologies from leading manufacturers across the industry. We are committed to quality and service, delivering strong customer satisfaction. We stock over 10,000 pre-owned lab equipment, components, and mix-and-match parts for your systems.

Select the fields to be shown. Others will be hidden. Drag and drop to rearrange the order.
  • Image
  • SKU
  • Rating
  • Price
  • Stock
  • Availability
  • Add to cart
  • Description
  • Content
  • Weight
  • Dimensions
  • Additional information
Click outside to hide the comparison bar
Compare