Shimadzu TOC-L CPH Analyzer with Total Nitrogen Unit TNM-L & ASI-L Autosampler
Item ID:
Manufacturer
Condition
3-12 month warranty
Service options & contracts available
To sell, trade-in, or service your equipment
How do we Refurbish Lab Equipment?
Looking for further assistance? Contact Us
Request Quote/ Offer
Estimate Shipping Time
Canada & U.S.: 3-5 business days
International: 5-10 business days
Tariff-Free Shipping to the U.S.!
The Shimadzu TOC-L series of analyzers adopts the 680°C combustion catalytic oxidation method, which was developed by Shimadzu and is now used worldwide. While providing an ultra wide range of 4 μg/L to 30,000 mg/L, this analyzer boast a detection limit of 4 μg/L through coordination with NDIR. This is the highest level of detection sensitivity available with the combustion catalytic oxidation method. In addition, the combustion catalytic oxidation method makes it possible to efficiently oxidize not only easily-decomposed, low-molecular-weight organic compounds, but also hard-to-decompose insoluble and macromolecular organic compounds.
Includes:
- Shimadzu TOC-L CPH Total Organic Carbon Analyzer
- Shimadzu TNM-L Total Nitrogen Unit
- Shimadzu ASI-L Autosampler
- Necessary tubes and supplies
- PC and Software
Specifications:
| Type | Specifications |
|---|---|
| Measured Items | TC (total carbon)
IC (inorganic carbon) TOC (total organic carbon) NPOC (Non-Purgeable Organic Carbon) |
| Measurement Range | TC:50 μg/L – 30,000 mg/L
IC:4 μg/L – 3,000mg/L TN:20 μg/L – 10,000 mg/L |
| Carrier gas | High-purity air (CO, CO2, HC content: Each 1 ppm max., dew point: -50 °C max.) Supply pressure: 200 ± 10 kPa (Additional use of optional carrier gas regulator: 300 to 600 kPa) Optional use of nitrogen gas (not possible in the TN measurement). With the standard model, optional use of pressurized gas |
| Reproducibility | TC, NPOC: CV 1.5 % max. or ±50 µg/L max
IC: CV 1.5 % max. or ±4 µg/L max |
| Measuring time | TC: approx. 3 min
IC: approx. 4 min |
| Sample dilution | Dilution rate of 2x to 50x (automatic sample dilution by syringe pump), dilution accuracy: ± 2 % max. (2x to 20x), ± 5% max. (21x to 50x) |
| TOC-L | |
|---|---|
| Type | Specifications |
| Measurement range | 4 µg/L to 30,000 mg/L (TC) |
| TNM-L | |
|---|---|
| Type | Specifications |
| Measurement | Chemiluminescence |
| Measurement Range | up to 10000 mg/L |
| Detection limit | 5 µg/L (CPH, CSH)
20 µg/L (CPN, CSN) Reproducibility: CV 3 % max. |
| Measuring time | approx. 4 min |
| Ozone source gas | air (compressed or housing air) 500 mL/min |
| ASI-L | |
|---|---|
| Type | Specifications |
| Vial types | 9 mL, 24 mL, 40 mL |
| Number of vials | 9 mL: 93
24 mL: 93 40 mL: 68 |
People Also View
Related products
-
Shimadzu TOC-L CPN Analyzer with Total Nitrogen Unit TNM-L & ASI-L Autosampler
Item ID: RefurbishedIn Stock
SpectraLab Scientific Inc
Established in 2003,
SpectraLab Scientific is a global leader of Refurbished analytical equipment. We specialize in reconditioning equipment, integrating the best available technologies from leading manufacturers across the industry. We are committed to quality and service, delivering strong customer satisfaction. We stock over 10,000 pre-owned lab equipment, components, and mix-and-match parts for your systems.
Our Services
Spectralab Scientific Inc. offers a variety of laboratory services to help you get started. Whether you’re setting up a new labs or developing your start up, we can help you overcome obstacles and provide turn-key solutions.
Repair Service
Parts and repairs for analytical equipment
Qualifications
IQ/OQ/PQ centered on compliance and data integrity
Service Contracts
Protect your lab and minimize downtime
Warranty
Up to 5 years warranty on most of the equipment
Installation & Training
Hassle-free setup & training from our professionals
Maintenance
Yearly PM to ensure peak performance