The Shimadzu TOC-L series of analyzers adopts the 680°C combustion catalytic oxidation method, which was developed by Shimadzu and is now used worldwide. While providing an ultra wide range of 4 μg/L to 30,000 mg/L, this analyzer boast a detection limit of 4 μg/L through coordination with NDIR. This is the highest level of detection sensitivity available with the combustion catalytic oxidation method. In addition, the combustion catalytic oxidation method makes it possible to efficiently oxidize not only easily-decomposed, low-molecular-weight organic compounds, but also hard-to-decompose insoluble and macromolecular organic compounds.

Includes:

  • Shimadzu TOC-L CPH Total Organic Carbon Analyzer
  • Shimadzu TNM-L Total Nitrogen Unit
  • Shimadzu ASI-L Autosampler
  • Necessary tubes and supplies
  • PC and Software

Specifications:

Type Specifications
Measured Items TC (total carbon)

IC (inorganic carbon)

TOC (total organic carbon)

NPOC (Non-Purgeable Organic Carbon)

Measurement Range TC:50 μg/L – 30,000 mg/L

IC:4 μg/L – 3,000mg/L

TN:20 μg/L – 10,000 mg/L

Carrier gas High-purity air (CO, CO2, HC content: Each 1 ppm max., dew point: -50 °C max.) Supply pressure: 200 ± 10 kPa (Additional use of optional carrier gas regulator: 300 to 600 kPa) Optional use of nitrogen gas (not possible in the TN measurement). With the standard model, optional use of pressurized gas
Reproducibility TC, NPOC: CV 1.5 % max. or ±50 µg/L max

IC: CV 1.5 % max. or ±4 µg/L max

Measuring time TC: approx. 3 min

IC: approx. 4 min

Sample dilution Dilution rate of 2x to 50x (automatic sample dilution by syringe pump), dilution accuracy: ± 2 % max. (2x to 20x), ± 5% max. (21x to 50x)
TOC-L
Type Specifications
Measurement range 4 µg/L to 30,000 mg/L (TC)
TNM-L
Type Specifications
Measurement Chemiluminescence
Measurement Range up to 10000 mg/L
Detection limit 5 µg/L (CPH, CSH)

20 µg/L (CPN, CSN)

Reproducibility: CV 3 % max.

Measuring time approx. 4 min
Ozone source gas air (compressed or housing air) 500 mL/min
ASI-L
Type Specifications
Vial types 9 mL, 24 mL, 40 mL
Number of vials 9 mL: 93

24 mL: 93

40 mL: 68

SpectraLab Scientific Inc
Established in 2003,

SpectraLab Scientific is a global leader of Refurbished analytical equipment. We specialize in reconditioning equipment, integrating the best available technologies from leading manufacturers across the industry. We are committed to quality and service, delivering strong customer satisfaction. We stock over 10,000 pre-owned lab equipment, components, and mix-and-match parts for your systems.

Select the fields to be shown. Others will be hidden. Drag and drop to rearrange the order.
  • Image
  • SKU
  • Rating
  • Price
  • Stock
  • Availability
  • Add to cart
  • Description
  • Content
  • Weight
  • Dimensions
  • Additional information
Click outside to hide the comparison bar
Compare