Waters 486 Tunable Absorbance Detector
3-12 month warranty
Service options & contracts available
To sell, trade-in, or service your equipment
How do we Refurbish Lab Equipment?
Looking for further assistance? Contact Us
Request Quote/ Offer
Estimate Shipping Time
Canada & U.S.: 3-5 business days
International: 5-10 business days
Tariff-Free Shipping to the U.S.!
The Waters 486 tunable absorbance detector is a single-Channel tunable ultraviolet/visible (UV/VIS) detector designed for high performance liquid chromatography (HPLC) applications.
The Waters 486 tunable absorbance detector can operate as a stand-alone unit (with a chart recorder) or with other Waters system modules, such as the Waters 860 Networking Computer System, or the Water 745/745B or 746 integrators.
The waters 486 tunable absorbance detector provides a usable light intensity over the wavelength range of 190-600 nanometers (nm), Although its primary use is in the ultraviolet wavelength range between 190 and 380 nanometers. The maximum sensitivity of the instrument is 0.001 absorbance units full scale (AUFS).
Wavelength range and sensitivity
The Waters 486 tunable absorbance detector provides a usable light intensity over the wavelength range 190 to 600 nanometers (nm), although its primary use is in the ultraviolet range between 190 and 380 nm. The maximum sensitivity of the instrument is 0.001 absorbance units full scale (AUFS).
The Waters 486 tunable absorbance detector offers the following features:
- Easy to use operator interface
- Self-calibration to ensure wavelength accuracy
- Battery backup to retain parameter settings when the detector is powered down or during power interruptions
- A deuterium arc lamp
- The patented Taper-Cell flow cell design, which reduces interference caused by refractive index effects.
- Optional cells:
- Non-metallic
- Semi-prep
- Microbore
- High-pressure (optimized for operation in LC/MS) Lamp shut-off to lengthen usable lamp life
- User-selectable chart mark on IV integrator output (standard on 10 mV)
- Wavelength scanning to determine optimal wavelength
- Diagnostic routines to verify proper functioning
- A modular design for easy servicing
Operational Specifications
- Wavelength range (single, fixed wavelength): 190 to 600 nm
- Wavelength accuracy: ± 2 nm
- Wavelength repeatability: ± 0.25 nm
- Noise: ± 1.0 x 10-5 AU, 1 second filter setting, 210 to 280 nm) dry cell
- Drift:
- 1.0 x 10-4 AU per °C (after warm-up)
- 1.0 x 10-4 AU per hour (after warm-up)
- Sensitivity Setting Range: 0.001 to 2.0 AUFS (under software control)
- Filter Setting Range: 0.1 to 5.0 seconds (under software control)
- Bandwidth: 8 nm
Optical Component Specifications
- Lamp Source: Deuterium arc lamp
- Flow cell: Dual beam Taper-Cell design
- Path length: 10 mm (standard, analytical)
- Cell volume: 8 ul (standard, analytical)
- Pressure limit: 1000 psi
- Materials: 316 stainless steel, Dynasil Suprasil II, Teflon
Environmental Specifications
- Operating temperature: 15 to 32.2 °C
- Operating humidity: 20 to 80%, non-condensing
- Shipping and storage temperature: -40° to 65 °C
- Shipping and storage humidity: 20 to 80% non-condensing
This refurbished Waters 486 tunable absorbance detector is really in great condition and is tested at SpectraLab Scientific.
A 90 day warranty is included. Multiple available.
People Also View
Related products
SpectraLab Scientific Inc
Established in 2003,
SpectraLab Scientific is a global leader of Refurbished analytical equipment. We specialize in reconditioning equipment, integrating the best available technologies from leading manufacturers across the industry. We are committed to quality and service, delivering strong customer satisfaction. We stock over 10,000 pre-owned lab equipment, components, and mix-and-match parts for your systems.
Our Services
Spectralab Scientific Inc. offers a variety of laboratory services to help you get started. Whether you’re setting up a new labs or developing your start up, we can help you overcome obstacles and provide turn-key solutions.
Repair Service
Parts and repairs for analytical equipment
Qualifications
IQ/OQ/PQ centered on compliance and data integrity
Service Contracts
Protect your lab and minimize downtime
Warranty
Up to 5 years warranty on most of the equipment
Installation & Training
Hassle-free setup & training from our professionals
Maintenance
Yearly PM to ensure peak performance