This HP / Agilent 1200 Series HPLC system comes with the followings:
- G1322A Degasser
- G1310A Isocratic Pump
- G1329A Autosampler or Manual Injector
- G1316A TCC
- G1314B Variable Wavelength Detector
- Solvent Tray
- PC with Suitable software and all essential cables required for full operation
This Agilent 1200 Series HPLC system is in a great shape. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.
Performance Specifications
| G1310A Iso Pump | |
|---|---|
|
Hydraulic system |
Dual piston in series pump with proprietary servo-controlled variable stroke drive, floating pistons and active inlet valve |
|
Settable flow range |
0.001 – 10 ml/min, in 0.001 ml/min increments |
|
Flow range |
0.2 – 10.0 ml/min |
|
Recommended pH range |
1.0 – 12.5, solvents with pH > 2.3 should not contain acids which attack stainless steel. |
|
Operating voltage |
230V |
|
Hydraulic system |
Dual piston in series pump with proprietary servo-controlled variable stroke drive, floating pistons and active inlet valve |
|
Settable flow range |
0.001 – 10 ml/min, in 0.001 ml/min increments |
| G1329A Autosampler | |
|---|---|
| Type | Specification |
| GLP features | Early maintenance feedback (EMF), electronic records of maintenance and errors |
| Communications | Controller-area network (CAN). GPIB (IEEE-448), RS232C, APG-remote standard, optional four external contact closures and BCD vial number output |
| Safety features | Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
| Injection range | 0.1 – 100 µl in 0.1 µl increments Up to 1500 µl with multiple draw (hardware modification required) |
| Replicate injections | 1 – 99 from one vial |
| Precision | Typically < 0.5 % RSD of peak areas from 5 – 100 µl, Typically < 1 % RSD of peak areas from 1 – 5 µl |
| Minimum sample volume | 1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in 300 µl microvial |
| Carryover | Typically < 0.1 %, < 0.05 % with external needle cleaning |
| Sample viscosity range | 0.2 – 50 cp |
| Replicate injections per vial | 1 – 99 |
| Sample capacity | 100 × 2-ml vials in 1 tray
40 × 2-ml vials in ½ tray 15 × 6-ml vials in ½ tray (Agilent vials only) |
| Injection cycle time | Typically 50 s depending on draw speed and injection volume |
| G1316A TCC | |
|---|---|
| Type | Specification |
| Temperature range | 10 degrees below ambient to 80 °C
up to 80 °C: flow rates up to 5 ml/min |
| Temperature stability | ± 0.15 °C |
| Temperature accuracy | ± 0.8 °C
± 0.5 °C (with calibration) |
| Column capacity | Three 30 cm |
| Warm-up/cool-down time | 5 minutes from ambient to 40 °C
10 minutes from 40 – 20 °C |
| Dead volume | 3 µl left heat exchanger
6 µl right heat exchanger |
| Communications | Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN via other 1200 series module |
| Safety and maintenance | Extensive diagnostics, error detection and display (through control module and Agilent ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas. |
| GLP features | Column-identification module for GLP
documentation of column type |
| Housing | All materials recyclable |
| G1314B VWD | |
|---|---|
|
Type |
Specification |
|
Detection type |
Double-beam photometer |
|
Light source |
Deuterium lamp |
|
Noise |
± 0.75 x 10 -5 AU, at 254 nm |
|
Drift |
3 x 10-4AU/hr, at 254 nm |
|
Linearity |
>2 AU upper limit |
|
Wavelength range |
190-600 nm |
|
Wavelength accuracy |
± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter |
|
Wavelength bandwidth |
6.5 nm typical over whole wavelength range |
|
Maximum data rate |
13 Hz |
|
Flow cells |
Standard: 14-µL volume, 10-mm cell path length and 40 bar (588 psi) pressure maximum High Pressure: 14-µL volume, 10-mm cell path length and 400 bar (5880 psi) Pressure maximum Micro: 1-µL volume, 5-mm cell path length and 40 bar (588 psi) pressure maximum Semimicro: 5-µL volume, 6-mm cell path length and 40 bar (588 psi) pressure maximum |
|
Communications |
Controller-area network (CAN), RS-232. APG Remote:ready, start, stop and shut-down signals, LAN optional |
|
Safety and maintenance |
Extensive diagnostics, error detection and display through control module and Agilent ChemStation, leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas. |
|
GLP |
Electronic records of maintenance and errors. Early maintenance feedback (EMF) for continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded. Verification of wavelength accuracy with built-in holmium oxide filter. |
|
Housing |
All materials recyclable |
People Also View
Related products
-
HP Agilent 1100 Radio-HPLC system with IN/US Beta-Ram Radio-HPLC Detector
Item ID: RefurbishedIn Stock -
HP 1050 Quat Pump, HP 1050 VWD and HP 1050 Thermostatted column compartment
Item ID: RefurbishedIn StockPrice:$4300 USD
SpectraLab Scientific Inc
Established in 2003,
SpectraLab Scientific is a global leader of Refurbished analytical equipment. We specialize in reconditioning equipment, integrating the best available technologies from leading manufacturers across the industry. We are committed to quality and service, delivering strong customer satisfaction. We stock over 10,000 pre-owned lab equipment, components, and mix-and-match parts for your systems.
Our Services
Spectralab Scientific Inc. offers a variety of laboratory services to help you get started. Whether you’re setting up a new labs or developing your start up, we can help you overcome obstacles and provide turn-key solutions.
Repair Service
Parts and repairs for analytical equipment
Qualifications
IQ/OQ/PQ centered on compliance and data integrity
Service Contracts
Protect your lab and minimize downtime
Warranty
Up to 5 years warranty on most of the equipment
Installation & Training
Hassle-free setup & training from our professionals
Maintenance
Yearly PM to ensure peak performance