The Agilent 1260 Infinity G1377A High Performance Micro Autosampler is designed to perform capillary LC with injection of sample volumes ranging from nL to μL. A micro Rheodyne valve and the optimized design of the needle seat, loop and seat capillaries minimize dispersion. A high-resolution metering device offers resolution ten times better than a standard autosampler, bypass operation facilitates low delay volume, increased sample injection speed for high sample throughput, flexible and convenient sample handling with different types of sample containers. Using 384-well plates allows to process up to 768 samples unattended.

S/N: DEACP00171

Specifications:

Type Specifications
GLP features Early maintenance feedback (EMF), electronic records of maintenance and errors
Communications Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
Safety features Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
Injection range 0.01– 8 μl in 0.01 μl increments with the small loop capillary 0.01– 40 μl in 0.01 μl increments with the extended loop capillary
Precision Typically < 0.5% RSD of peak areas from 5 – 40 μl, Typically < 1% RSD from 1 – 5 μl Typically < 3% RSD from 0.2 – 1 μl
Pressure range up to 400 bar (5880 psi)
Sample viscosity range 0.2 – 5 cp
Sample capacity 2 × well-plates (MTP) + 10 × 2 ml vials

108 x 2-mL vials in 2 x 54 vial plate plus 10 additional 2 mL vials

30 x 6-mL vials in 2 x 15 vial plate plus 10 additional 2 mL vials

54 Eppendorf tubes (0.5/1.5/2.0 mL) in 2 x 27 Eppendorf tube plate

Injection cycle time Typically < 30 s using following standard conditions:

Default draw speed: 4 μl/min

Default eject speed: 10 μl/min

Injection volume: 0.1 μl

Carry-over Typically < 0.05% using the following conditions:

Column: 150 x 0.5 mm Hypersil ODS, 3 μm

Mobile phase: Water/Acetonitrile = 85/15

Column Flow rate: 13 μl/min

Injection volume: 1 μl caffeine (=25 ng caffeine), 1 μl water to test carryover

Outside wash of needle before injection: 20 sec with water using flush port

Dimensions 200mm x354mm x440mm

SpectraLab Scientific Inc
Established in 2003,

SpectraLab Scientific is a global leader of Refurbished analytical equipment. We specialize in reconditioning equipment, integrating the best available technologies from leading manufacturers across the industry. We are committed to quality and service, delivering strong customer satisfaction. We stock over 10,000 pre-owned lab equipment, components, and mix-and-match parts for your systems.

Select the fields to be shown. Others will be hidden. Drag and drop to rearrange the order.
  • Image
  • SKU
  • Rating
  • Price
  • Stock
  • Availability
  • Add to cart
  • Description
  • Content
  • Weight
  • Dimensions
  • Additional information
Click outside to hide the comparison bar
Compare